Datasheet SHT1x
Sensor Performance
Relative Humidity123
Parameter
Condition
Resolution 1
Accuracy 2
SHT10
typical
maximal
Accuracy 2
SHT11
typical
maximal
Accuracy 2
SHT15
typical
maximal
Repeatability
Hysteresis
Non-linearity linearized
Response time 3 τ (63%)
Operating Range
Long term drift 4 normal
min typ max Units
0.4 0.05 0.05 %RH
8 12 12 bit
±4.5
%RH
see Figure 2
±3.0
%RH
see Figure 2
±2.0
%RH
see Figure 2
±0.1
%RH
±1
%RH
<<1
%RH
8
s
0
100 %RH
< 0.5
%RH/yr
Temperature45
Parameter
Condition
Resolution 1
Accuracy 2
SHT10
Accuracy 2
SHT11
Accuracy 2
SHT15
Repeatability
typical
maximal
typical
maximal
typical
maximal
Operating Range
Response Time 6 τ (63%)
Long term drift
min typ max Units
0.04 0.01 0.01 °C
12 14 14 bit
±0.5
°C
see Figure 3
±0.4
°C
see Figure 3
±0.3
°C
see Figure 3
±0.1
°C
-40
123.8 °C
-40
254.9 °F
5
30 s
< 0.04
°C/yr
± 10
±8
±6
SHT10
±4
SHT11
±2
SHT15
±0
0
10 20 30 40 50 60 70 80 90 100
Relative Humidity (%RH)
Figure 2: Maximal RH-tolerance at 25°C per sensor type.
± 3.0
± 2.5
± 2.0
SHT10
± 1.5
± 1.0
± 0.5
± 0.0
-40 -20
SHT15
0 20 40 60
Temperature (°C)
Figure 3: Maximal T-tolerance per sensor type.
SHT11
80 100
Electrical and General Items
Parameter
Source Voltage
Power
Consumption 5
Communication
Storage
Condition min typ max Units
2.4 3.3 5.5 V
sleep
2 5 µW
measuring
3
mW
average
90
µW
digital 2-wire interface, see Communication
10 – 50°C (0 – 125°C peak), 20 – 60%RH
Packaging Information
Sensor Type
SHT10
SHT11
SHT15
Packaging
Tape & Reel
Tape & Reel
Tape & Reel
Tape & Reel
Tape & Reel
Tape & Reel
Quantity
2000
100
400
2000
100
400
Order Number
1-100218-04
1-100051-04
1-100098-04
1-100524-04
1-100085-04
1-100093-04
This datasheet is subject to change and may be
amended without prior notice.
1 The default measurement resolution of is 14bit for temperature and 12bit for
humidity. It can be reduced to 12/8bit by command to status register.
2 Accuracies are tested at Outgoing Quality Control at 25°C (77°F) and 3.3V.
Values exclude hysteresis and are applicable to non-condensing environments
only.
3 Time for reaching 63% of a step function, valid at 25°C and 1 m/s airflow.
4 Value may be higher in environments with high contents of volatile organic
compounds. See Section 1.3 of Users Guide.
5 Values for VDD=3.3V at 25°C, average value at one 12bit measurement
per second.
6 Response time depends on heat capacity of and thermal resistance to
sensor substrate.
www.sensirion.com
Version 5 – December 2011
2/12